GTM43004P-WWVV-x.x-P3 (SR with NA Plug 3 pin) Potted/Encapsulated in plastic housing Medical Power Supply Datasheet
Request Information
Download PDF
Configure Output Connector
GTM43004P-WWVV-x.x-P3 (SR with NA Plug 3 pin) Image | |||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Model Image | ![]() | ||||||||||||||||||||||||||||||||||
Information | |||||||||||||||||||||||||||||||||||
Model Number | GTM43004P-WWVV-x.x-P3 (SR with NA Plug 3 pin) | ||||||||||||||||||||||||||||||||||
Description | GTM43004P-WWVV-x.x-P3 (SR with NA Plug 3 pin), Medical Power Supply, Potted/Encapsulated in plastic housing, Regulated Switchmode AC-DC Power Supply AC Adaptor, , Input Rating: 100-240V~, 50-60 Hz, North America Blades with Ground Prong, Class I, (3) Conductors, Output Rating: 120 Watts, Power rating with convection cooling (W) , 5-48V in 0.1V increments, Approvals: cRUus; WEEE; China RoHS; CE; VCCI; Ukraine; RoHS; PSE; Level V; Designed to meet EN 60950-1; Designed to meet EN 60601-1; Designed to meet 60335; cULus; Designed to meet 60950-1; Class I; | ||||||||||||||||||||||||||||||||||
Agency Documents | http://www.globtek.info/certs/GTM43004P/ | ||||||||||||||||||||||||||||||||||
CE EC-Declaration for Medical | Contact GlobTek for more information | ||||||||||||||||||||||||||||||||||
CE EC-Declaration for Industrial/Commercial/ITE | https://globtek.secure.force.com/models/ec_declaration?id=a0Oa000000OJ9VwEAL | ||||||||||||||||||||||||||||||||||
CE EC-Declaration for Household | Contact GlobTek for more information | ||||||||||||||||||||||||||||||||||
UKCA-Declaration for Medical | Contact GlobTek for more information | ||||||||||||||||||||||||||||||||||
UKCA-Declaration for Industrial/Commercial/ITE | https://globtek.secure.force.com/models/ukca_declaration?id=a0Oa000000OJ9VwEAL | ||||||||||||||||||||||||||||||||||
UKCA-Declaration for Household | Contact GlobTek for more information | ||||||||||||||||||||||||||||||||||
Morocco SDoC Declaration | https://globtek.secure.force.com/models/Morocco_SDoC_Declaration?id=a0Oa000000OJ9VwEAL | ||||||||||||||||||||||||||||||||||
RoHS/RoHS3 Declaration | https://www.globtek.com/pdf/rohs_cert/a0Oa000000OJ9VwEAL | ||||||||||||||||||||||||||||||||||
Prop 65 Statement | https://en.globtek.com/prop-65-statement.pdf | ||||||||||||||||||||||||||||||||||
REACH Declaration | https://www.globtek.com/pdf/iso_certificates/REACH.pdf | ||||||||||||||||||||||||||||||||||
Conflict Minerals Declaration | https://www.globtek.com/pdf/conflict-minerals.pdf | ||||||||||||||||||||||||||||||||||
Model Parameters | |||||||||||||||||||||||||||||||||||
Type | Potted/Encapsulated in plastic housing / vergossenes Gerät im Plastikgehäuse | ||||||||||||||||||||||||||||||||||
Technology | Regulated Switchmode AC-DC Power Supply AC Adaptor | ||||||||||||||||||||||||||||||||||
Category | Medical Power Supply / Netzgerät Medizinisch | ||||||||||||||||||||||||||||||||||
Input Voltage Range (V) | 100-240V~, 50-60 Hz | ||||||||||||||||||||||||||||||||||
Max Input Current (A) | 2.0A | ||||||||||||||||||||||||||||||||||
Max Output Power Rating (W) | 120 | ||||||||||||||||||||||||||||||||||
Output Voltage Range (V) | 5-48 | ||||||||||||||||||||||||||||||||||
Efficiency | V | ||||||||||||||||||||||||||||||||||
Ingress Protection | IP68 | ||||||||||||||||||||||||||||||||||
Size (mm) | 178.0*76.0*39.5 | ||||||||||||||||||||||||||||||||||
Enclosure | |||||||||||||||||||||||||||||||||||
Rating Table | |||||||||||||||||||||||||||||||||||
* (Final ordering #/Part No to be defined)
| |||||||||||||||||||||||||||||||||||
Specifications | |||||||||||||||||||||||||||||||||||
A) ELECTRICAL SPECIFICATIONS:
01. Input Voltage: Specified 90-264 Vac, Nameplate rated: 100-240Vac 02. Input Frequency: Specified 47-63 Hz, Nameplate rated 50-60Hz 03. Output Regulation: +/- 5% measured at the output connector 04. Line Voltage Regulation: +/- 1% typical measured at the output connector 05. Output Ripple (Vp-p): +/-1% or 200 mV whichever is greater, measured at 20 MHz bandwidth with 0.1 uf ceramic capacitor in parallel with 47 uf electrolytic capacitor connected at the end of the output connector at nominal line 06. Turn-ON/OFF Overshoot: 10% maximum, 1 mS typical recovery time for 25% step load 07. Turn-ON Delay: 3000 mS, maximum @ full load, nominal line 08. Hold-Up Time: 10 mS typical @ nominal input voltage and full load 09. Inrush Current: 40A typical 115VAC input and 60A typical at 230VAC input 10. Switching Frequency: 100 KHz typical B) PROTECTION 01. Over-Voltage: AC cycling required to reset output 02. Short Circuit: AC cycling required to reset output 03. Input Protection: Input line fusing C) SAFETY 1. Dielectric Withstand Voltage: 5656 Vdc from primary to secondary 2. Earth Leakage Current: Maximum allowed values: 5 mA NC(Normal condition); 10 mA SFC(single fault condition) 3. Touch Current: Maximum allowed values: 100uA NC(Normal condition) 500uA SFC(single fault condition) 4. Earth continuity test:<0.1 Ohm or < 0.2 Ohm when with a non-detachable Power Supply cord 5. Means of Protection: 2 x MOPP 6. Primary to Secondary Bridging Capacitor: (2) * Y1 type capacitors in series D) OTHER: 01. MTBF: 200,000 Hours @ 25°C ambient temperature 02. Operating Temperature: 0°C to 40°C ambient temperature 03. Humidity: 0% to 90% relative humidity 04. Storage Temperature: -10°C to 80°C 05. Cooling: Convection E) ENCLOSURE 01. Size: 178.0*76.0*39.5 mm +/-1.0 mm 02. Markings: Label | |||||||||||||||||||||||||||||||||||
Derating Curve | |||||||||||||||||||||||||||||||||||
Qty (ea.)/Carton | 20 | ||||||||||||||||||||||||||||||||||
Individual Box/Gift Box Size (mm) | Egg Crated | ||||||||||||||||||||||||||||||||||
Individual Box/Gift Box Color | None | ||||||||||||||||||||||||||||||||||
Master Carton Size (mm) | 461 x 460 x 182 | ||||||||||||||||||||||||||||||||||
Gross Weight (kg) | 15.6 | ||||||||||||||||||||||||||||||||||
Label Width (mm) | 60.0 | ||||||||||||||||||||||||||||||||||
Label Height (mm) | 60.0 | ||||||||||||||||||||||||||||||||||
Input Configuration | |||||||||||||||||||||||||||||||||||
Description | North America Blades with Ground Prong, Class I, (3) Conductors | ||||||||||||||||||||||||||||||||||
Output Configuration | |||||||||||||||||||||||||||||||||||
Common output connector options: Click to add to your request.
For a comprehensive list of options, click here Contact GlobTek for your specific requirements or custom solutions. | |||||||||||||||||||||||||||||||||||
Approvals | |||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||
Popular Products (Stock items) | |||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||
News Releases | |||||||||||||||||||||||||||||||||||
|
Wie funktioniert ein AC/DC-Adapter?
Der AC/DC-Adapter oder das externe Netzteil besteht aus einem Netzteil, das in einem Gehäuse verbaut und zum Anschluss an das Wechselstromnetz mit einem Stecker versehen ist. Deshalb wird häufig auch von einem Steckernetzteil gesprochen. Das Netzteil wird in die Steckdose gesteckt und die Wechselspannung wird dann vom AC/DC-Adapter in eine Gleichspannung umgewandelt, die über das angeschlossene Ausgangskabel in das zu versorgende Gerät eingespeist wird.