GTM964500PWWWVV.V-F Open Frame/Internal ICT / ITE / Medical Power Supply Datasheet

GTM964500PWWWVV.V-F Image
Model ImageGTM964500PWWWVV.V-F, ICT / ITE / Medical Power Supply, Open Frame/Internal, Regulated Switchmode AC-DC Power Supply AC Adaptor, Forced airflow for full rated power (CFM): 10CFM, Input Rating: 100-240V~, 50-60 Hz, , Output Rating: 450 Watts, Power rating with convection cooling (W) 250W, 12-54V in 0.1V increments, Approvals: GTM964500PWWWVV.V-F, ICT / ITE / Medical Power Supply, Open Frame/Internal, Regulated Switchmode AC-DC Power Supply AC Adaptor, Forced airflow for full rated power (CFM): 10CFM, Input Rating: 100-240V~, 50-60 Hz, , Output Rating: 450 Watts, Power rating with convection cooling (W) 250W, 12-54V in 0.1V increments, Approvals:
Information
Model NumberGTM964500PWWWVV.V-F
DescriptionGTM964500PWWWVV.V-F, ICT / ITE / Medical Power Supply, Open Frame/Internal, Regulated Switchmode AC-DC Power Supply AC Adaptor, Forced airflow for full rated power (CFM): 10CFM, Input Rating: 100-240V~, 50-60 Hz, , Output Rating: 450 Watts, Power rating with convection cooling (W) 250W, 12-54V in 0.1V increments, Approvals: GTM964500PWWWVV.V-F, ICT / ITE / Medical Power Supply, Open Frame/Internal, Regulated Switchmode AC-DC Power Supply AC Adaptor, Forced airflow for full rated power (CFM): 10CFM, Input Rating: 100-240V~, 50-60 Hz, , Output Rating: 450 Watts, Power rating with convection cooling (W) 250W, 12-54V in 0.1V increments, Approvals:
Agency Documents
CE EC-Declaration for Medicalhttps://globtek.secure.force.com/models/ec_declaration?id=a0O3a00000Ly9m5EAB
CE EC-Declaration for Industrial/Commercial/ITEhttps://globtek.secure.force.com/models/ec_declaration?id=a0O3a00000Ly9m5EAB
CE EC-Declaration for HouseholdContact GlobTek for more information
UKCA-Declaration for Medicalhttps://globtek.secure.force.com/models/ukca_declaration?id=a0O3a00000Ly9m5EAB
UKCA-Declaration for Industrial/Commercial/ITEhttps://globtek.secure.force.com/models/ukca_declaration?id=a0O3a00000Ly9m5EAB
UKCA-Declaration for HouseholdContact GlobTek for more information
Morocco SDoC Declarationhttps://globtek.secure.force.com/models/Morocco_SDoC_Declaration?id=a0O3a00000Ly9m5EAB
RoHS/RoHS3 Declarationhttps://www.globtek.com/pdf/rohs_cert/a0O3a00000Ly9m5EAB
Prop 65 Statementhttps://en.globtek.com/prop-65-statement.pdf
REACH Declarationhttps://www.globtek.com/pdf/iso_certificates/REACH.pdf
Conflict Minerals Declarationhttps://www.globtek.com/pdf/conflict-minerals.pdf
Model Parameters
TypeOpen Frame/Internal / internes /offenes
TechnologyRegulated Switchmode AC-DC Power Supply AC Adaptor
CategoryICT / ITE / Medical Power Supply / ICT / ITE / Medical Power Supply
Input Voltage Range (V)100-240V~, 50-60 Hz
Max Input Current (A)6A
Max Output Power Rating (W)450
Output Voltage Range (V)12-54
Auxiliary/Fan Output Voltage (V)12
Auxiliary/Fan Output Power (W)15W Forced Air, 10W Convection
EfficiencyUSA DOE Level VI / Eco-design Directive 2009/125/EC, (EU) 2019/1782
Ingress ProtectionN/A
Size (mm)127.0 (L) x 76.2 (W) x 38.5 (H)
EnclosureUser-added image

Customer may need to provide EMI shielding if not in metal chassis
User-added image
Rating Table
Model Number (Final ordering #/Part No to be defined)VoltageAmps (Convection)Amps (Forced Air)Watts (Convection)Watts (Forced Air)RFQ
* (Final ordering #/Part No to be defined)
Specifications

Features

  • High power density 3" x 5"� format
  • Auxiliary +12V fan output
  • Auxiliary +5V standby output
  • Power-good, remote-off, & remote-sense
  • Class I, Class II, or Class II with Functional Earth input configurations
  • ITE/ICT and medical (2 x MOPP) certified

Input

Input Voltage:Specified: 90-264VAC, Nameplate: 100-240VAC
Input Frequency:Specified: 47-63Hz, Nameplate: 50-60Hz
Input Current:5.8A typ. @ 90VAC, 2.3A typ. @ 230VAC (full load)
No Load Input Power:<150mW
Inrush Current:<35A @ @ 100VAC , <70A @ 230VAC (cold start @ 25°C)
Efficiency:85 - 95% typ. (full load)
Power Factor:>0.95 @ 115VAC, >0.90 @ 230VAC (full load)

Output

Output Power:For 12-18.9V output: 400W (25CFM forced air), 225W (convection), inclusive of fan/aux power
For 19-54.0V output: 450W (25CFM forced air), 250W (convection), inclusive of fan/aux power
Turn-on Delay:< 3 seconds @ 90VAC, < 2 seconds @ 115VAC
Output Voltage Tolerance:± 2.0% (worst case, set point + load/line regulation)
Load Regulation:± 0.5% (measured at connector)
Line Regulation:± 0.2% (measured at connector)
Ripple:1% pk-pk (using a 47µF low-ESR cap + 0.1µF ceramic capacitor, measured @ 20MHz BW)
Transient Response:<5% deviation (with 25% load step), recovery to within 1% in 1ms
Startup Overshoot:<5% (full load)
Hold-up Time:8ms typ. (full load)
+12V Fan Output:800mA (max)
+5V Standby Output:1A (max)

Protections

Input Protection:MOV transient suppressor, dual fuses (line + neutral)
Over-Voltage Protection:110-150%. Latched-off, cycle AC to reset
Over-Current Protection:105-150% typ. Non-latched with auto-recovery
Short Circuit Protection:Non-latched with auto-recovery
Over-Temperature Protection:Latched-off, cycle AC to reset

Environmental

MTBF:200,000 hours @ 25°C ambient
Operating Temperature:-20°C to 40°C (Full output power), -40°C to 80°C (Derated)
Storage Temperature:-40°C to 85°C
Humidity:0% to 95% relative humidity, non-condensing
Altitude:5000m max.
Vibration:Conforms to EN60068-2-6

Safety

Dielectric Withstand Voltage:4000VAC or 5656VDC from input to output
3000VAC or 4242VDC from input to earth
Intended Medical Application:Type-B or Type-BF
Touch Current:NC: <100µA SFC: <500µA
Earth Leakage CurrentNC: <0.5mA SFC: <1.0mA
Means of Protection:2 x MOPP
Primary-Secondary Bridging Capacitance:Two IEC60384-14 Y1 capacitors in series

EMC

Applicable Standards:Medical: EN60601-1-2, 4th edition
Emissions: EN55032, EN 55011, EN61000-6-3, EN61000-6-4, CISPR11/CISPR22
Immunity: EN55024, EN61000-6-1, EN61000-6-2
Conducted Emissions:Class B, FCC Part 15, Class B (with resistive load)
Radiated Emissions:Class B, FCC Part 15, Class B (with resistive load)
Harmonic Current Voltage Distortion:EN61000-3-2, Class A
Voltage Fluctuations/Flicker:EN61000-3-3
Electrostatic Discharge (ESD) Immunity:EN61000-4-2, 10KV contact Discharge, 20KV air discharge (applied to input or output terminals)
Radiated RF Immunity:EN61000-4-3, 10V/m 80-2700MHz, 80% 1KHz AM.
EFT/Burst Immunity:EN61000-4-4, 4KV/100kHz.
Line Surge Immunity:EN61000-4-5, 2KV differential, 4KV common-mode
Conducted RF Immunity:EN61000-4-6, 3VRMS, 80% 1KHz AM
Power Frequency Magnetic Field Immunity:EN61000-4-8, 30A/m
Voltage Dip Immunity:EN61000-4-11

 

Derating CurveUser-added image
User-added image
Qty (ea.)/Carton
Individual Box/Gift Box Size (mm)
Individual Box/Gift Box Color
Master Carton Size (mm)TBD
Gross Weight (kg)
Label Width (mm)TBD
Label Height (mm)TBD
Input Configuration
DescriptionUser-added image
Optional Input Wiring Kits are available:
PN/Ordering NoLengthEnd AEnd B
C810070M6VAR2(R)60IEC 60320 C8JST VAR-2
C1410070M6VAR2(R)60IEC 60320 C13JST VAR-2
VAR2210070M7VAR2(R)73JST VAR-2JST VAR-2
Custom lengths, construction, cable types, ferrites, filters, and termination available upon request.
Approvals

Below follows a link to the latest regulatory certifications and marks for this specific product. National standards as well as certification schemes are constantly changing and so is the set of certification marks and symbols applicable to this product. Various standards, marks and symbols, and certificates are subject to change at any time without notice.

Additional standards or unique country certifications are available upon request. Please contact Globtek to request a quotation.

LogoDescription
Conforms to UL STD. 62368-1
Certified to CSA STD C22.2 NO.62368-1
Conforms to UL STD. 62368-1 Certified to CSA STD C22.2 NO.62368-1
User-added image
 
Morocco SDoC declaration http://www.globtek.info/certs/Morocco%20SDoC%20Declaration/
User-added imageUKCA Certification
Popular Products (Stock items)
SKUVolts OutCurrent Out (Amps)ConnectorRFQSample Request

Was ist ein Open-Frame-Netzteil zum Einbau?

Open-Frame-Netzteile werden auch als AC/DC-Wandler bezeichnet. Es handelt sich dabei um ein Netzteil zum Einbau. Das Open-Frame-Netzteil wird in das Innere eines Systems eingebaut und dient zur Umwandlung des Wechselstroms in Gleichspannung.

Was ist ein Netzteil nach IEC 62368-1:2018?

Die IEC 62368-1: 2018 gilt für die Sicherheit elektrischer und elektronischer Geräte im Bereich der Audio-, Video-, Informations- und Kommunikationstechnik sowie für den Geschäfts- und Bürobereich mit Nennspannungen von bis zu 600 V. Es handelt sich um einen Produktsicherheitsstandard, der Energiequellen klassifiziert, Schutzmaßnahmen hinsichtlich dieser Energiequellen vorschreibt und Leitlinien für die Anwendung dieser Schutzmaßnahmen sowie die Anforderungen daran darlegt. Die vorgeschriebenen Schutzmaßnahmen sollen die Wahrscheinlichkeit des Auftretens von Schmerz, Verletzungen und Sachschäden im Brandfall verringern. Ein AC/DC-Wandler oder Netzteil nach IEC 62368-1 ist ein Netzteil, das nach dieser Norm geprüft und zertifiziert wurde.


Haben IEC 62368-1-Netzteile die IEC60950-1-Netzteile abgelöst?

Ab dem 20. Dezember 2020 werden die Sicherheitsstandards IEC 60065 und IEC 60950-1, die bislang für Geräte der Informations-, Audio- und Videotechnik galten, von IEC 62368-1 abgelöst. In Nordamerika gelten die ETL- und UL 60950-1-Zertifizierungen weiterhin und können bis auf Weiteres weiterhin verwendet werden.


Was ist der Unterschied zwischen einem Netzteil nach IEC 62368-1 und IEC 60950-1?

Während die Struktur und der Inhalt der Normen im Allgemeinen sehr unterschiedlich sind, entsprechen die meisten Produkte, die zuvor der IEC 60950-1 entsprachen, der neuesten Norm IEC 62368-1 und es sind keine oder nur geringe Änderungen erforderlich.


Ist mein Netzteil nach IEC60950-1 weiterhin verfügbar?

GlobTek wird die Netzteile nach IEC 60950-1 bis auf weiteres auf unbestimmte Zeit herstellen und liefern.

Was ist eine Effizienz nach Level VI?

Die Stufen I bis VI sind Teil eines internationales Protokolls zur Effizienzkennzeichnung, mit dem Hersteller die Effizienz eines Produkts bestimmen können. Diese Leistung basiert auf der Messung der Effizienz im aktiven Modus und der Leistung im Leerlaufmodus von externen Netzteilen oder AC/DC-Adaptern, die bei verschiedenen Ausgangsleistungen getestet werden. Die Leistung im Leerlaufmodus bezieht sich auf den Stromverbrauch, den ein externes Netzteil auch dann aufweist, wenn die Anwendung ausgeschaltet oder getrennt wird.


Ist eine Effizienz nach Level VI erforderlich?

Eine Effizienz nach Level VI ist in bestimmten Ländern und Regionen für bestimmte Produkte oder bestimmte Anwendungen vorgeschrieben.


Was ist die Ökodesign-Richtlinie 2009/125/EG?

Mit der Ökodesign-Richtlinie wurden in Europa die neuesten Anforderungen an Energieeffizienz und Energieeinsparung für AC/DC-Adapter und externe Netzteile erlassen.

Wie funktioniert ein AC/DC-Adapter?

Der AC/DC-Adapter oder das externe Netzteil besteht aus einem Netzteil, das in einem Gehäuse verbaut und zum Anschluss an das Wechselstromnetz mit einem Stecker versehen ist. Deshalb wird häufig auch von einem Steckernetzteil gesprochen. Das Netzteil wird in die Steckdose gesteckt und die Wechselspannung wird dann vom AC/DC-Adapter in eine Gleichspannung umgewandelt, die über das angeschlossene Ausgangskabel in das zu versorgende Gerät eingespeist wird.

Information anfordern

Model

Call
Email
💬
Text
💁
Chat
?
Angebotsanfrage
Können wir Ihnen helfen?
Können wir Ihnen helfen?